Magnetic Resonance Imaging on Patients with Implanted Cardiac Pacemakers

Seitz, Sebastian

Abstract:
The aim of this work was to identify the patterns that can induce heating around implanted cardiac pacemakers during MRI and to develop strategies to counteract them. Two approaches were taken: computer simulations of the occurring electromagnetic field distributions and in-vitro experiments using phantoms in real MRI devices, both for conventional bore-hole and new open MRI systems. Using the open MRI, the observed heating could be reduced significantly.


Zugehörige Institution(en) am KIT Institut für Biomedizinische Technik (IBT)
Publikationstyp Hochschulschrift
Jahr 2011
Sprache Englisch
Identifikator ISBN: 978-3-86644-610-6
ISSN: 1864-5933
URN: urn:nbn:de:0072-211877
KITopen ID: 1000021187
Verlag KIT Scientific Publishing, Karlsruhe
Umfang 156 S.
Serie Karlsruhe transactions on biomedical engineering / Ed.: Universität Karlsruhe (TH) / Institute of Biomedical Engineering ; 10
Abschlussart Dissertation
Fakultät Fakultät für Elektrotechnik und Informationstechnik (ETIT)
Institut Institut für Biomedizinische Technik (IBT)
Prüfungsdaten 14.10.2010
Referent/Betreuer Prof. O. Dössel
Schlagworte MRI, pacemaker, numerical field calculation, biomedical engineering