KIT | KIT-Bibliothek | Impressum | Datenschutz

Multimodal eye's optical quality (MEOQ)

Garcia Guerra, Carlos Enrique

Abstract:
This document summarizes the work done during the implementation of the Multimodal Eye's Optical Quality (MEOQ) system, a measurement device that integrates a double-pass (DP) instrument and a Hartmann-Shack (HS) sensor. The system has been used to quantify scattering in the human eye. Moreover, the configurable properties of the system have been used to explore a method for reducing speckle in systems that rely on reflections of light in the ocular fundus.

Open Access Logo


Volltext §
DOI: 10.5445/IR/1000063014
Cover der Publikation
Zugehörige Institution(en) am KIT Institut für Technik der Informationsverarbeitung (ITIV)
Publikationstyp Hochschulschrift
Publikationsjahr 2016
Sprache Englisch
Identifikator urn:nbn:de:swb:90-630144
KITopen-ID: 1000063014
Verlag KIT, Karlsruhe
Umfang XXI, 158 S.
Abschlussart Dissertation
Fakultät Fakultät für Elektrotechnik und Informationstechnik (ETIT)
Institut Institut für Technik der Informationsverarbeitung (ITIV)
Prüfungsdaten 29.07.2016
Referent/Betreuer Prof. W. Stork
KIT – Die Forschungsuniversität in der Helmholtz-Gemeinschaft
KITopen Landing Page