KIT | KIT-Bibliothek | Impressum | Datenschutz

Reflection on boon and bane of Water Absorbing Components in Active Implants during Package Testing and Operation

Koker, Liane ORCID iD icon 1; Gengenbach, Ulrich 1; Bretthauer, Georg 1
1 Institut für Automation und angewandte Informatik (IAI), Karlsruher Institut für Technologie (KIT)

Abstract:

For package testing of active implants internal humidity measurements are performed. Thereby, the water absorbency of the measurement components can affect the measurement result, leading to an erroneously prolonged predicted lifetime. Here, the influence of internal components is discussed using an exemplary setup. It is described how to model and consider the components’ absorbencies for lifetime estimation and how beneficial they are during implant operation.


Verlagsausgabe §
DOI: 10.5445/IR/1000088251
Veröffentlicht am 04.12.2018
Cover der Publikation
Zugehörige Institution(en) am KIT Institut für Automation und angewandte Informatik (IAI)
Publikationstyp Zeitschriftenaufsatz
Publikationsmonat/-jahr 09.2018
Sprache Englisch
Identifikator ISSN: 2364-5504
urn:nbn:de:swb:90-882516
KITopen-ID: 1000088251
HGF-Programm 43.22.03 (POF III, LK 01) Printed Materials and Systems
Erschienen in Current directions in biomedical engineering
Verlag De Gruyter
Band 4
Heft 1
Seiten 319–322
Schlagwörter Humidity measurement; accelerated aging; lifetime; packaging; implant; hermeticity
Nachgewiesen in Dimensions
Scopus
KIT – Die Forschungsuniversität in der Helmholtz-Gemeinschaft
KITopen Landing Page