Microfluidic Impedance Biosensor Chips Using Sensing Layers Based on DNA-Based Self-Assembled Monolayers for Label-Free Detection of Proteins

Alsabbagh, Khaled 1; Hornung, Tim 1; Voigt, Achim 1; Sadir, Sahba 2; Rajabi, Taleieh 1; Länge, Kerstin 1
1 Institut für Mikrostrukturtechnik (IMT), Karlsruher Institut für Technologie (KIT)
2 Institut für Mikroverfahrenstechnik (IMVT), Karlsruher Institut für Technologie (KIT)

Abstract:

A microfluidic chip for electrochemical impedance spectroscopy (EIS) is presented as bio-sensor for label-free detection of proteins by using the example of cardiac troponin I. Troponin I is one of the most specific diagnostic serum biomarkers for myocardial infarction. The microfluidic impedance biosensor chip presented here consists of a microscope glass slide serving as base plate, sputtered electrodes, and a polydimethylsiloxane (PDMS) microchannel. Electrode functionalization protocols were developed considering a possible charge transfer through the sensing layer, in addition to analyte-specific binding by corresponding antibodies and reduction of nonspecific protein adsorption to prevent false-positive signals. Reagents tested for self-assembled monolayers (SAMs) on gold electrodes included thiolated hydrocarbons and thiolated oligonucleotides, where SAMs based on the latter showed a better performance. The corresponding antibody was covalently coupled on the SAM using carbodiimide chemistry. Sampling and measurement took only a few minutes. Application of a human serum albumin (HSA) sample, 1000 ng/mL, led to negligible impedance changes, while application of a troponin I sample, 1 ng/mL, led to a significant shift in the Nyquist plot. ... mehr


Verlagsausgabe §
DOI: 10.5445/IR/1000131675
Veröffentlicht am 23.04.2021
Originalveröffentlichung
DOI: 10.3390/bios11030080
Scopus
Zitationen: 14
Dimensions
Zitationen: 18
Cover der Publikation
Zugehörige Institution(en) am KIT Institut für Mikrostrukturtechnik (IMT)
Institut für Mikroverfahrenstechnik (IMVT)
Publikationstyp Zeitschriftenaufsatz
Publikationsjahr 2021
Sprache Englisch
Identifikator ISSN: 2079-6374
KITopen-ID: 1000131675
HGF-Programm 43.35.02 (POF IV, LK 01) Functionality of Soft Matter and Biomolecular Systems
Erschienen in Biosensors
Verlag MDPI
Band 11
Heft 3
Bemerkung zur Veröffentlichung Gefördert durch den KIT-Publikationsfonds
Schlagwörter biosensor; immunosensor; cardiac troponin I; single-strand DNA; electrochemical impedance spectroscopy; label-free; proteins; microfluidic chip; self-assembled monolayers