KIT | KIT-Bibliothek | Impressum | Datenschutz

Integration of maXs-type microcalorimeter detectors for high-resolution x-ray spectroscopy into the experimental environment at the CRYRING@ESR electron cooler

Pfäfflein, Ph ; Allgeier, S.; Bernitt, S.; Fleischmann, A.; Friedrich, M.; Hahn, Ch; Hengstler, D.; Herdrich, M. O.; Kalinin, A.; Kröger, F. M.; Kuntz, P.; Lestinsky, M.; Löher, B.; Menz, E. B.; Over, T.; Spillmann, U.; Weber, G.; Zhu, B.; Enss, Ch 1; ... mehr

Abstract:

We report on the first integration of novel magnetic microcalorimeter detectors (MMCs), developed within SPARC (Stored Particles Atomic Physics Research Collaboration), into the experimental environment of storage rings at GSI$^6$, Darmstadt, namely at the electron cooler of CRYRING@ESR. Two of these detector systems were positioned at the 0° and 180° view ports of the cooler section to obtain high-resolution x-ray spectra originating from a stored beam of hydrogen-like uranium interacting with the cooler electrons. While previous test measurements with microcalorimeters at the accelerator facility of GSI were conducted in the mode of well-established stand-alone operation, for the present experiment we implemented several notable modifications to exploit the full potential of this type of detector for precision x-ray spectroscopy of stored heavy ions. Among these are a new readout system compatible with the multi branch system data acquisition platform of GSI, the synchronization of a quasi-continuous energy calibration with the operation cycle of the accelerator facility, as well as the first exploitation of the maXs detectors' time resolution to apply coincidence conditions for the detection of photons and charge-changed ions.


Verlagsausgabe §
DOI: 10.5445/IR/1000152008
Veröffentlicht am 27.10.2022
Originalveröffentlichung
DOI: 10.1088/1402-4896/ac93be
Scopus
Zitationen: 10
Dimensions
Zitationen: 8
Cover der Publikation
Zugehörige Institution(en) am KIT Institut für Prozessdatenverarbeitung und Elektronik (IPE)
Publikationstyp Zeitschriftenaufsatz
Publikationsjahr 2022
Sprache Englisch
Identifikator ISSN: 0031-8949, 1402-4896
KITopen-ID: 1000152008
HGF-Programm 54.12.03 (POF IV, LK 01) Science Systems
Erschienen in Physica Scripta
Verlag Royal Swedish Academy of Sciences
Band 97
Heft 11
Seiten Art.-Nr.: 114005
Vorab online veröffentlicht am 14.10.2022
Schlagwörter microcalorimeter, precision x-ray spectroscopy, heavy highly charged ions, ion storage ring, metallic magnetic calorimeter, coincidence measurement
Nachgewiesen in Web of Science
Dimensions
Scopus
Relationen in KITopen
KIT – Die Forschungsuniversität in der Helmholtz-Gemeinschaft
KITopen Landing Page