KIT | KIT-Bibliothek | Impressum | Datenschutz

Active Eye Lens Dosimetry With Dosepix: Influence of Measurement Position and Lead Glass Shielding

Ullmann, Leonie ; Beißer, Florian; Behrens, Rolf; Funk, Stefan; Hilgers, Gerhard; Hupe, Oliver; Roth, Jürgen; Tröltzsch, Tom 1; Zutz, Hayo; Miche, Thilo
1 Institut für Meteorologie und Klimaforschung Atmosphärische Spurengase und Fernerkundung (IMKASF), Karlsruher Institut für Technologie (KIT)

Abstract:

In this work, the effect of the measurement position on measurements of Hp(3) of a new active eye lens dosemeter prototype (ELD) based on the Dosepix detector is examined. A comparison between measuring directly in front of the eye and measuring at the side of the head of an Alderson phantom showed no significant influence on the resulting Hp(3) for different radiation qualities and angles. In addition, to account for the absorption effect of radiation safety glasses, pieces of lead glass were attached to the front and side of the dosemeter. The challenging geometry of the human face, when additionally equipped with absorbing glasses and irradiated at different angles of radiation incidence, led to a complex irradiation setup, whereby the dosemeter cannot be placed at the real point of interest, i.e., directly in front of the eye. Corresponding effects and consequences for radiation protection measurements have been investigated by using a human like Alderson head phantom as well as thermoluminescent dosemeters (TLDs) and the active ELD based on the Dosepix detector. For specific angles, the radiation bypassed the radiation safety glasses and lead glass pieces, leading to an increase in the measured Hp(3). ... mehr


Zugehörige Institution(en) am KIT Institut für Meteorologie und Klimaforschung Atmosphärische Spurengase und Fernerkundung (IMKASF)
Publikationstyp Zeitschriftenaufsatz
Publikationsmonat/-jahr 04.2026
Sprache Englisch
Identifikator ISSN: 2469-7311, 2469-7303
KITopen-ID: 1000186702
Erschienen in IEEE Transactions on Radiation and Plasma Medical Sciences
Verlag IEEE Canada
Band 10
Heft 4
Seiten 619–628
Vorab online veröffentlicht am 16.09.2025
Schlagwörter Active personal dosemeter, Dosepix, dosimetry for interventional procedures, dosimetry for radiation-based medical applications, eye lens dosimetry, radiation detectors for medical applications: semiconductors