KIT | KIT-Bibliothek | Impressum | Datenschutz

Calibration and evaluation of MET models for estimating energy expenditure using thigh and ankle-worn move 4 accelerometer

Giurgiu, Marco 1; Hysenllari, Eni; Ottenbacher, Jörg; Ebner-Priemer, Ulrich 1; Timm, Irina ORCID iD icon 1
1 Institut für Sport und Sportwissenschaft (IfSS), Karlsruher Institut für Technologie (KIT)

Abstract:

$\textbf{Objectives}$
Valid estimation of energy expenditure remains a challenge, particularly when using ankle- and thigh-worn devices. The Move 4 is a research-grade accelerometer previously tested for predicting metabolic equivalents (METs) when worn at the waist or wrist. This study aimed to calibrate and evaluate regression models to estimate METs from Move 4 data when worn at the ankle and thigh.

$\textbf{Methods}$
Participants completed walking and jogging tasks under laboratory conditions while wearing Move 4 sensors and with indirect calorimetry as a reference measure. Models were calibrated using study 1 (n=160) and evaluated in an independent dataset (study 2; n=15). Performance was assessed using mean absolute error (MAE), root mean square error (RMSE), and Bland-Altman analyses.

$\textbf{Results}$
The MET models demonstrated strong agreement across both locations and datasets. For the thigh position, the MAE ranged from 0.60 METs (walking) to 1.38 METs (jogging), with RMSE of 0.82 and 1.70 in the evaluation data. Calibration metrics were comparable (jogging: MAE=1.24, RMSE=1.63). The ankle models showed similar accuracy, with MAEs of 0.66 (walking) and 1.39 (jogging), and RMSEs of 0.85 and 1.67, respectively. ... mehr


Verlagsausgabe §
DOI: 10.5445/IR/1000193733
Veröffentlicht am 01.06.2026
Originalveröffentlichung
DOI: 10.1515/teb-2025-0027
Cover der Publikation
Zugehörige Institution(en) am KIT Institut für Sport und Sportwissenschaft (IfSS)
Publikationstyp Zeitschriftenaufsatz
Publikationsdatum 04.12.2025
Sprache Englisch
Identifikator ISSN: 2942-6812
KITopen-ID: 1000193733
Erschienen in Translational exercise biomedicine
Verlag De Gruyter
Band 2
Heft 4
Seiten 315–325
Schlagwörter validation; evaluation; accelerometer; sensor; energy expenditure; model performance