KIT | KIT-Bibliothek | Impressum

Nanotomography endstation at the P05 beamline : Status and perspectives

Greving, I.; Ogurreck, M.; Marschall, F.; Last, A.; Wilde, F.; Dose, T.; Burmester, H.; Lottermoser, L.; Müller, M.; David, C.; Beckmann, F.

Abstract:
The Imaging Beamline IBL/P05 at the DESY storage ring PETRA III, operated by the Helmholtz-Zentrum Geesthacht, has two dedicated endstations optimized for micro- and nanotomography experiments [1-3]. Here we present the status of the nanotomography endstation, highlight the latest instrumentation upgrades and present first experimental results. In particular in materials science, where structures with ceramics or metallic materials are of interest, X-ray energies of 15 keV and above are required even for sample sizes of several 10 μm in diameter. The P05 imaging beamline is dedicated to materials science and is designed to allow for imaging applications with X-ray energies of 10 to 50 keV. In addition to the full field X-ray microscopy setup, the layout of the nanotomography endstation allows switching to cone-beam configuration. Kinematics for X-ray optics like compound refractive lenses (CRLs),
Fresnel zone plates (FZP) or beam-shaping optics are implemented and the installation of a Kirkpatrick Baez-mirror (KB mirror) system is foreseen at a later stage of the beamline development. Altogether this leads to a high flexibility of ... mehr


Zugehörige Institution(en) am KIT Institut für Mikrostrukturtechnik (IMT)
Publikationstyp Zeitschriftenaufsatz
Jahr 2017
Sprache Englisch
Identifikator DOI: 10.1088/1742-6596/849/1/012056
ISSN: 1742-6588, 1742-6596
URN: urn:nbn:de:swb:90-722234
KITopen ID: 1000072223
Erschienen in Journal of physics / Conference Series
Band 849
Heft 1
Seiten Art.Nr.: 012056
Lizenz CC BY 3.0 DE: Creative Commons Namensnennung 3.0 Deutschland
KIT – Die Forschungsuniversität in der Helmholtz-Gemeinschaft KITopen Landing Page