The New Generation of the KIT 3D USCT

Gemmeke, Hartmut; Berger, L.; Hopp, Torsten ORCID iD icon; Zapf, Michael; Tan, W.; Blanco, R.; Leys, R. ORCID iD icon; Peric, I. ORCID iD icon; Ruiter, Nicole V. ORCID iD icon

Abstract:

The first clinical studies with our current prototype, 3D USCT II, enabled us to identify the necessary improvements for transition of our method to clinical practice. The main goals are to improve the contrast of reflection and transmission tomography, and to optimize the coverage of the imaged breast by a new geometry of the transducer distribution. Furthermore, for cost-effective industrial mass production, a self-calibration method allows us to relax the precision of the positioning of the transducers to 0.1 mm. The readout of the transducer arrays is now carried out by an ASIC, developed for a more cost-effective design. The coupling of the measuring device to the patient was optimized to cover the full size of the breast up to the pectoral muscles. Finally, the data acquisition and readout time were reduced to 1.5 minutes each by new micro-TCA electronics and larger FPGAs.


Verlagsausgabe §
DOI: 10.5445/IR/1000079770
Veröffentlicht am 15.02.2018
Cover der Publikation
Zugehörige Institution(en) am KIT Institut für Prozessdatenverarbeitung und Elektronik (IPE)
Publikationstyp Proceedingsbeitrag
Publikationsdatum 30.01.2018
Sprache Englisch
Identifikator ISBN: 978-3-7315-0689-8
urn:nbn:de:swb:90-797704
KITopen-ID: 1000079770
HGF-Programm 54.02.02 (POF III, LK 01) Ultraschnelle Datenauswertung
Erschienen in Proceedings of the International Workshop on Medical Ultrasound Tomography: 1.- 3. Nov. 2017, Speyer, Germany. Hrsg.: T. Hopp
Veranstaltung 1st International Workshop on Medical Ultrasound Tomography (MUST 2017), Speyer, Deutschland, 01.11.2017 – 03.11.2017
Verlag KIT Scientific Publishing
Seiten 271-282
Relationen in KITopen
KIT – Die Forschungsuniversität in der Helmholtz-Gemeinschaft
KITopen Landing Page